C-650 Operation Manual, Version B
5 Putting into operation
18
5 Putting into operation
This chapter describes how the device is installed and gives instructions on initial startup.
NOTE
Inspect the device for damages during unpacking. If necessary, prepare a status report immedi-
ately to inform the postal company, railway company or transportation company.
Keep the original packaging for future transportation.
5.1 Installation site
Put the device on a stable, horizontal surface. Consider the maximum product dimensions and
weight. Obtain the environmental conditions as described in section 3.2 Technical data overview.
Installation prerequisites:
Make sure, the device is not subject to drafts or significant temperature changes.
Do not place it near air conditioning vents, windows, ovens, etc.
Make sure that access to the power supply to disconnect the device must be kept at any
time.
Make sure to leave a distance of at least 20cm between the device and another object or
wall.
Place the detector close to the outlet of the column to minimize extra column band broaden-
ing which would reduce the resolution of the choromatographic separation.
WARNING
Death or serious injuries by use in explosive environments.
Do not operate the device in explosive environments
Do not operate the device with explosive gas mixtures
Before operation, check all gas connections for correct installation
Directly withdraw released gases and gaseous substances by sufficient ventilation
CAUTION
Risk of minor or moderate injury by heavy weight of the device.
Consult a second person to lift the device
Do not drop the device
Place the device on a stable, even and vibration-free surface
Keep limbs out of crushing zone
To transport the device, place your hands under the device
NOTICE
Risk of device damage by liquids or mechanical shocks.
Do not spill liquids over the device or its components
Do not move the device when it is loaded with sample liquid
Keep external vibrations away from the device
Safely attach the device to the bench in earthquake prone regions
NOTE
As a destructive detector, the C-650 should be the last one in the flow path or can be used with
a split ter.