7 Maintenance and repairs
47 K-415 Operation Manual, Version B
Fig. 7.2: Flushing the pump
7.9 Hose cooling control / condenser
Check the hose monthly and replace it as soon as it becomes porous.
7.10 Hose water control feed
Check the sieve once a month for residues and clean it if necessary.
Check the adapter, its sealings and the sieve on a yearly base and replace defective parts.
7.11 Power supply fuses
1
2
To replace the power supply fuses, proceed as
follows:
Pull out the power supply cable.
Remove the fuse holder a.
Replace the defective fuse with a new fuse b of
the same type:
100 V: T2A L 250 V
120 V: T2A L 250 V
230 V: T1A L 250 V
Put the fuse holder back.
Fig. 7.3: Replacing the power supply fuses