6 Operation
59 K-439 Operation Manual, Version A
8.3 Device fuses
To change a defective fuse proceed as follows
Switch off the SpeedDigester and unplug the power cord.•
To access the glass fuses on the rear side use a flat screwdriver blade. •
Turn it approx. 5 turns anticlockwise to free the insert.
Take out the insert together with the fuse.
Replace the defective fuse (see technical data for fuse type!)•
Reconnect the SpeedDigester to the mains.•
!
WARNING
Death or serious burns by electric current.
Switch off the instrument•
Disconnect the power cord and prevent unintentional restart before removing the fuse holder•
Do not touch the fuse or the fuse holder with wet hands•
Exchange defective fuses only with the original type•
T 10A L 250 V
Note
Device fuses may blow occasionally due to voltage peaks under high system load. When fuses burn
frequently, inform your customer service!