6 Operation
24 B-440 Operation Manual, Version D
6.3 Setting up application parameters
Internal program storage
The B-440 is capable of storing 9 application methods in its program storage.s
Each program has 5 steps with individual parameter pairs of time and temperature.s
Parameter name Range
Programs 1 – 9
Steps 1 – 5
Temperature ambient – 600 °C
Time (per "STEP") 0 – 240 minutes
Setting up PROGRAMs and STEPs
Switch "ON" the Wet Digester. s
๗After the bootprocess ( 4 seconds) the instrument will be operational.
Select a program via the right "SELECT" button (below "PROGRAM" display).s
Press the left "SELECT" button (below "STEP" display) repeatingly to switch through every s
"TIME" / "TEMP" parameter of each "STEP" in a loop.
๗The "STEP" display shows the selected step (1—5 ).
๗The corresponding "TIME" / "TEMP" display will flash for 10 seconds and switch to "STEP" level 1
without button action.
To change either the heating time or temperature of a step, use the "UP" and "DOWN" buttons s
while the respective display flashes.
๗As soon as the display stops flashing, the value is being stored.
NOTE
Some application methods do not require a full set of 5 steps. To disable a "STEP", set its "TIME " s
to 0.
During a running application, programming is not possible.s
6.4 How to perform a wet application
Digestion fumes can be hazardous and must be neutralized by a Scrubber or a condensate
trap / Scrubber combination. Fumes which may escape during operation must be safely withdrawn by
a fume hood! Wear personal protective equipment, see section 2.5 for general safety information!
!
WARNING
Death or serious poisoning by contact or incorporation of harmful substances and fumes at use.
Before operation, check the instrument for correct assemblings
Rinse condensate trap, hoses and connectors before uses
Before operation inspect sealing surfaces and tubes for good condition and tightnesss
Secure all hoses with hose clampss
Exchange worn out or defective parts immediatelys
Only operate the instrument in a fume hoods
Directly withdraw released gases and gaseous substances by a Scrubbers
Dispose of condensate in tubings and glassware safely after uses
Do not inhale digestion fumess