Buchi PrepChrom C-700 Instrukcja Użytkownika Strona 49

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 48
BÜCHILabortechnikAG Installation|5
PrepChromC700 49/110
u MovetheupperholderdownontothecolumnandtightenthePEEK
adapter(5).
u Tightentheupperknurledknob(1).
5.6 Mixingchambervolume,adjustment
NOTE
Theadjustmentofthemixingchamberisrecommendedtominimizethedead
volume.
u Turnthewheel(2)toadjustthemixingchamber(3)accordingtothede
siredflowrateofthepump.
Turningthewheelclockwisewillincreasethemixingvalue.
Turningthewheelcounterclockwisewilldecreasethemixingvalue.
u Readthecurrentvalueonthescaleoftheprotrudingpiston(1).
NOTE
Thepistoncanbeturnedwitha11/32"spannerifthescaleisturnedtowards
thedevice.
Flowrate Mixingchamber
0–80mL/min 0%
80–130mL/min 40%
130–175mL/min 80%
>175mL/min 100%
Przeglądanie stron 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 109 110

Komentarze do niniejszej Instrukcji

Brak uwag