BÜCHILabortechnikAG Operation|6
PrepChromC700 79/110
u Selectthedesiredsignal(1)forthechannel1–4orusethescanmode.
u Adjustthewavelength(nm)(2).
Additionalwavelengthscanbechosenforthechannels2–4.
5.7.11 Detection step, activation/deactivation [} page65]
u ChooseYes(3)inCollectifyouwanttomonitorthecollectionatthecho
senwavelength.
u Selectthedesiredthreshold.
u SetfiltersforthedetectorinF1(4);1=nofilter,10=maximumfilter
Info
Additionalcollectionstepscanbeadded.5.7.12 Collection step, activation/de‐
activation [} page65]
u SelectthetabCollection.
u Selectthedesiredtime(1).
u SelectthelocalminimumandchoosebetweenYesorNo(2).
u Selectthedesiredtubevolume(3).
u SelectthemodeThresholdtocollectelutedpeaks(4)orselectthemode
AlltocollectallorselectthemodeWaste.
NOTE
ThecolumnActionisinactiveduringmethodprogramming.Afterarun,itcon
sistsofallinrunmodificationswhichweremadeinordertobeabletorepro
duceit.