8 Troubleshooting
145
K-375/376/377 Operation Manual, Version B
Problem Cause Corrective measure
Nitrogen content too low Incomplete digestion
Insufficient H
2SO4
Kjeldahl Tablets and H2SO4 in wrong
ratio
Nitrogen content per sample tube is
too high
Not enough NaOH or incorrect
concentration of NaOH used (required
is 32 %)
Leakage during distillation
Leakage during digestion
Wrong titrant used
Defective pH electrode
Dirty glassware
Increase digestion time
Increase volume
Correct ratio
Not more than 200 mg Nitrogen per
sample tube
Correct volume until color change is
visible
Check and tighten, check connec-
tion between condenser and splash
protector, replace seal if required
Check sealing and scrubber suction
capacity
Check and correct
Calibrate electrode, replace if neces-
sary
Only use clean glassware
Poor repeatability Air bubbles in titration system, buret,
tubes
Aspiration not working properly
Incorrect or no pH electrode calibra-
tion
Sample inhomogeneous
Sample weighing problems
Incomplete digestion, digestion time
too short
Suction capacity during digestion is
too strong
Stirrer is not working
Dip tube blocked, loose, too short or
defective
Fix tubing and refill buret
Check for leaks and fix
Calibrate electrode with fresh buffer
Homogenize sample
Use weighing boats to improve
procedure
Check color of samples during
digestion and choose digestion time
accordingly
Reduce suction power at the
Scrubber with the bypass valve
Clean the stirrer, replace if necessary
Check and correct