8 Troubleshooting
148
K-375/376/377 Operation Manual, Version B
14’100 Titrator timeout Titration endpoint not reached. Check dosing unit,
if enough titration solution is available or update
frimware.
14’101 Titrator measured value is out of range Verify the buffers used for calibration, check the
electrode.
14’501 Dosing unit not ready, please check Verify if the dosing unit is connected to the distillation
unit.
14’502 Dosing unit locked Check the dosing unit and switch distillation unit off
and on.
14’503 Dosing unit not ready, no exchange unit Check if dosing unit is connected to the distillation
unit
14’504 Dosing unit not ready, no dosino Check if dosing unit is connected to the distillation
unit
14’505 Dosing unit overload Call service
14’506 Dosing unit not ready, cock blocked Dismantle the dosing unit (see chapter “Trouble-
shooting the dosing unit”)
14’602 Titrator stopped, maximal volume reached Make sure the electrode is in good working condi-
tions, the correct titration solution is used and that no
air bubbles are in titrator hoses.
14’603 Titrator stopped, stop endpoint reached Check the titrator and if enough titration solution is
available
14’604 Titrator stopped, stopPot reached Check the titrator and if enough titration solution is
available
14’605 Titrator stopped, stop time reached Check the titrator and if enough titration solution is
available
15’001 No sampler connected Switch on sampler and check the connection cable
15’002 Sampler: Target not achieved Try again or call service
15’003 Sampler: Connection lost Check the connection cable
15’101 Sampler: Shield open Close shield
15’102 Sampler: Crash detected Try again or call service
15’103 Sampler: Tube not found Put sample tube in place or call service
15’104 Sampler: Tube not released Remove sample tube or call service
15’105 Sampler: Error 5, reserve (gibt es nicht) Call service
15’106 Sampler: Error 6, reserve (gibt es nicht) Call service
15’107 Sampler: Shield not locked Call service
15’108 Sampler: Error reference position Call service
15’109 Sampler zeroadjustment not possible
because x or y deviation is larger then 3mm
or sampler was not in reference position
before adjustment start
Try again or call service
15’110 Sampler: Position error X-axis Call service
15’111 Sampler: Position error Y-axis Call service
15’112 Sampler: Position error Z-axis down Call service
15’113 Sampler: Position error Z-axis up Call service