5 Putting into operation
49
K-375/376/377 Operation Manual, Version B
On the K-375
• Remove the tightening screw from the holder
on the valve 1.
(The screw is not required for connection to
K-377)
• Slide the ring of the chain fastener onto the
holder on the valve 1 and fix it by tightening
the threaded bar.
• Slide the the plastic holder of the transfer
connection onto the threaded bar 2. Hold it in
place by screwing the second nut hand-tight
on top.
• Mount the white transfer hose on the valve
using the provided screw connection 3.
• Mount the red steam hose on the steam valve
and secure it with a hose clamp 4.
5.4 Reagent/water and waste connections
NOTICE
Risk of device damage by exceeding the maximum permissible pressure for the cooling water inlet.
• Make sure never to exceed the maximum permissible pressure of 6 bar for the cooling water
inlet.
!
WARNING
Serious chemical burns by corrosives.
• Make sure that the tanks are connected correctly. If the wrong tank (e. g. reagent tank
containing NaOH) is connected to the pump labelled as „H
2O”, NaOH might be dosed while
H2O is expected.