Buchi Multivapo P-12 Instrukcja Użytkownika Strona 41

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 40
5 Putting into operation
41 Multivapor™ Operation Manual, Version E
5.4 Glass assembly
NOTICE
Risk of glass breakage by excessive strains.
Mount all glassware parts without strains
Check glassware for proper fixing regularly and readjust fixing points if necessary
Do not use defective glassware
Use the protective shield (optional)
5.4.1 Type S and type C condenser
1
1
Install the condenser assemblies on the pro-
vided support or on a stable laboratory rod
ensuring that the stand base is oriented in the
direction of the condenser.
Secure the receiving flask with the clip a pro-
vided for this purpose.
Fig. 5.24: Installation of the condenser assembly on the support
5.4.2 Condenser assembly with the refrigerated receiver (optional)
2
3
1
The refrigerated receiver a can be used as an
alternative to the receiving flask and is secured
using the clip b. The internal cooling loop is
connected to a cooling source c (tap water or
recirculating chiller).
Fig. 5.25: Type S condenser with the refrigerated receiver