Rotavapor R-250
5 Operation
Fig. 5.1: Operating unit of the R-250
5 Operation
Make sure that the unit has been commissioned properly as
described in Chapter 4.
5.1 Layout of Operating and Display Elements
Main breaker switch
Lift for the bath
Rotation ON/OFF
Rotating speed adjustment
Rotating speed display
Heater ON/OFF
Input: Set-point temperature for bath
Temperature UP/DOWN
Display for bath temperature
Display for vapour temperature
Aeration
When the unit is switched off or in the event of a power failure,
the bath can be lowered automatically so that the evaporating
flask is always separated from the source of heat.
5.2 Setting the Maximum Setpoint Temperatu-
re and Lowering the Bath Automatically
Setting the Maximum Setpoint Temperature
s 0RESSTHEBUTTONh3%44%-0v, simultaneously swit-
ching on the main breaker switch . This activates the
input for setting the maximum set-point temperature.
s 4HEVALUECANBESETATANYLEVELDESIREDBYPRESSINGTHE
h$/7.vANDh50vBUTTONS.
s 4HEINPUTISSTOREDANDTHEUNITSWITCHEDINTOTHEOPERATING
MODEONLYAFTERTHEh!%2!4%vBUTTON is then pressed.