Buchi Syncor Polyvap Instrukcja Użytkownika Strona 15

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 46
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 14
Syncore Platform
15
4.4 Function test for drive and speed indica-
tor
There should be no objects of any kind resting upon the
base plate or between the base-plate and the housing of
the equipment.
• Setaspeedof60RPMbyturningthespeedcontroller
to the right.
• Thedisplayunit should indicate “60” and the base
plate of the basic unit should move slowly without any
pronounced vibration.
• Followingcompletionofthefunctiontest,resetthespeed
to 0 RPM
4.5 Function test heating and time
There should be no objects of any kind resting upon the
base plate or between the base-plate and the housing of
the equipment.
• Press once on the „UP“ button on the temperature
con trol unit. The display shifts from the current tempe-
ra ture to the set-point temperature.
• Pressthe„UP“buttonrepeatedlytosetanewset-point
for temperature of +40°C.
• After3seconds,theunittakesoverthisnewset-point.
• Pressthe„START“buttonrepeatedly.Thebaseplateis
heated to the set-point temperature. A dot in the tempe-
rature display shows that the heater has been turned
ON.
• Thetimedisplayshows how much time has passed
since starting.
If not, contact a local BUCHI service agent.
• TurntheheaterOFFafterthefunctionaltestbypressing
once on the “STOP” button.
• Tocheckthetemperatureoftheheatingplate,insertan
external thermometer (Ø 4 mm) into the reference socket.
4 Putting into Operation
Figure 9: Controlling the basic functions from the control panel
Przeglądanie stron 14