Buchi Syncor Polyvap Instrukcja Użytkownika Strona 20

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 46
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 19
Syncore Platform
20
5 Operation
5.4 Program mode, temperature/time steps
Typical application:
• Denitionofthetemperatureandtimeinseveralstagesat
a constant RPM, e.g. in a parallel evaporation at different
temperatures.
• Repetitionofheatingcyclesfor,e.g.materialtestingcycles
Schematic:
Programming mode, controlled parameters
1. Activating the programming function:
Press once on the „PROGRAM“ button.
The display indicates the first step (Step 1).
2. Entering the temperature for step 1:
Adjust the temperature set-point by pressing repeatedly
on the „UP“ or „DOWN“ button of the temperature control
.
3. Entering the time for step 1:
Press repeatedly on the „UP“ or „DOWN“ buttons of the
time control to define the time for step 1. The unit
will heat the base plate to this temperature and remains
constant for the time given.
4. Activating the next step:
Press once on the „PROGRAM“ button to activate
the next step in the program. The display indicates
the second step (step 2).
5. Entering the temperature and time for step 2:
Analogous to points 2 and 3.
6. Defining the last step in the program:
Press the „PROGRAM“ button once. The display
indicates the next step. Set the time to 00:00. This ter-
minates the prgram after the preceding step, i.e. heating
is switched off and the equipment continues to rotate
at a constant RPM or it stops (depending on the safety
settings).
Ifthe„UP“-Keyofthetimecontrollerisholduntil„OFF“
appears in the display (following the display of 9:59),
the equipment continues to operate at the temperature
denedinthisstepuntilthe„STOP“-Keyispressed.
Figure 14: Display of the basic unit after starting up the equipment
Przeglądanie stron 19