Syncore Platform
8
2 Safety
2.3 Authorized use
This unit has been designed and built as a laboratory
instru ment. The authorized application for the Syncore single
platform in con junction with accessories from the Syncore
product line is as follows:
Syncore Polyvap/Syncore single platform
with accessories:
• Parallelevaporation of solvents invarious formats and
con tainers across a range of temperatures from room
tempe ra ture up to + 150 °C and within a range of pres-
sures from 1 mbar up to ambient pressure.
Syncore Analyst:
• Parallelconcentrationofsolventsfrom6differentsamples
across a range of temperatures from room temperatures
up to + 100 ° (or bypassing pre-set safety settings, up to
150 °C), within a range of pressures from 1 mbar up to
ambient pressure.
Applications for the Syncore line of equipment
include:
• Parallelorganic-inorganicchemicalsynthesisintheliqui-
dor in the solid phase.
• Parallelmaterialinvestigations,digestions,preparationof
samples of from 4 to 96 samples, with sample volumes
ranging from 500 down to 1 ml.
• Parallelevaporationoffrom4to96sampleswithsample
volumes from 1 ml up to 500 ml after chemical synthesis,
after chromatography, after extraction, etc.
• Parallelconcentrationinchemicalanalysis,guaranteeing
a residual volume after the evaporation.
2.4 Unauthorized use
Any use other than those indicated above and any application
that does not correspond to the Technical Data is considered
to be improper use.
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment
may be impaired.
The operator himself bears sole respon si bi lity for any risks
arising from such use.
The following in par ti cu lar are explicitly forbidden:
• The processing of samples that can explode or ignite
explo si vely as the result of impact, friction, heat, or spar-
king (e.g., explosives, etc.).
• Thecarryingoutofchemicalreactionsthatcanproduce
sub stances that can explode or ignite explosively as the
re sultof impact, friction, heat, or sparking (e.g., explosives,
etc.);
• Use of the unit in rooms that require explosion-proof
equip ment;
Komentarze do niniejszej Instrukcji