5 Putting into operation
39 R-220 SE Operation Manual, Version F
Adjustment instructions
Make sure the foam sensor is installed correctly. Remove the white plastic screw a to get access
to the sensor adjustment screw.
Allow sufficient foam formation.
As soon as the foam touches the sensor housing, the LED at the sensor must switch active,
causing the aeration valve to open for an instant. The valve makes a noticable *click* sound!
If the sensor does not detect the foam use a small flat head screw driver to adjust the sensitivity
of the sensor.
Check, if the LED switches off as soon as the foam collapses. If not, slightly decrease the
sensitivity and start a new adjustment cycle.
a
b
NOTE
The foam sensor can only be used at decending glassware setups!