6|Operation BÜCHILabortechnikAG
24/42 OperationManualInterfaceI100
Fig.14:Displayduringevacuationsequence
u Toincreasethespecifiedpressurewhileevacuationisinprogress,presstheP↑
button(2).
Thesystempressureisincreasedbyasmallamountandthepumpswitchesto
Holdmode.Thecurrentpressureismaintainedevenifthespecifiedsettinghasnot
yetbeenreached.
u TocancelHoldmode,presstheHOffbutton(1).
Thespecifiedpressurepreviouslysetisreinstated.
u Tostopthedistillationprocess,presstheSTOPbutton.
u Ifnecessary,dependingontheventingsetting(seeChapter5.4.1"Controllerset
tings",page20),presstheSTOPbuttonagaintoventthesystem.
6.2.2 Continuousmode
u PresstheCont.button.
ThevacuumpumpisswitchedonandkeepsrunninguntiltheSTOPorMan.but
tonispressed.Afteranhour,thevacuumpumpswitchestoEcomode(seevac
uumpumpoperatinginstructions).