6|Operation BÜCHILabortechnikAG
26/42 OperationManualInterfaceI100
6.3 Calibratingthepressuresensor
ThepressuresensoriscalibratedatthefactorybyBUCHIpriortodelivery.However,it
canberecalibratedwiththeaidofanexternalreferencepressuregaugeatanytime.
u PresstheMenubuttonandselectPressuresensorcalibration.
Fig.16:Menu>Pressuresensorcalibration
6.3.1 Offsetcalibration
Anoffsethastobespecifiedifthepressureindicatedontheinterfaceunitdiffersfrom
thepressureindicatedbyanexternalreferencepressuregauge.Thedifferencebe
tweenthetwofiguresisreferredtoastheoffset.Theoffsetappliestotheentirepres
surerange.
u Onthecalibrationmenu,selecttheoptionCalibrationpressureoffset.
Fig.17:Menu>Pressuresensorcalibration>Calibrationpressureoffset
u Usingthearrowbuttons,settheoffsetandpressOKtoconfirm.
Themeasuredreadingswillthenautomaticallybeadjustedbythespecifiedoffset.
u Makesurethatthereferencepressureindicatedontheinterfaceunitmatchesthe
figureshownontheexternalreferencepressuregauge.