6.3.2 Simplecalibration(withouttemperaturecompensation)
Thiscalibrationmethodisusedtoadjustthelinearityofthepressuresensoratspecific
predefinedpressurelevels.Duringthecalibrationsequencethepressuresensoris
calibratedatroomtemperatureinsixstagesatnormalatmosphericpressureandat
800,600,400,200and10mbar.
NOTE
Thiscalibrationprocessmayonlybecarriedoutbyauthorizedservicetechnicians.
Duringthecalibrationsequence,thesystemcarriesoutaplausibilitycheckonthe
measuredreadingsandindicatesafaultiftheconfirmedreadingsareimplausible.
u Connectthereferencepressuregaugetothesamevacuumsystemthatthepump
andinterfaceunitareconnectedto.
u Onthecalibrationmenu,selecttheoptionSimplepressurecalibration.
Fig.18:Menu>Pressuresensorcalibration>Simplepressurecalibration
u Usingthearrowbuttons,adjustthespecifiedreferencepressuretothefigure
shownonthereferencepressuregaugeandpressOKtoconfirm.
u PresstheNextbutton.
u UsingtheP↑andP↓buttons,increase/decreasethevacuuminthesystemuntilthe
pressureshownonthereferencegaugeiswithin±10mbarofthespecifiedsetting.
u Repeatthelastthreestepsforeachofthepressurelevelslistedabove.
u Oncompletionofthelaststep,pressOKtoconfirmthecalibration.
Thecalibrationsettingsaresavedandthedisplayrevertstothedefaultscreen.