BÜCHILabortechnikAG Operation|6
OperationManualInterfaceI100 25/42
6.2.3 Timermode
InTimermodethesetpressureismaintainedforthesettime.Dependingonthevent
ingsetting(seeChapter5.4.1"Controllersettings",page20),thesystemistheneither
immediatelyventedoramessagesimplydisplayedindicatingthatdistillationhasfin
ished.
u PresstheMenubuttonandforModeselectTimer.
Fig.15:Mainmenu
u PresstheProg.button.
u Usingthearrowbuttons,setthepressureandthetimeandpressOKtoconfirmin
eachcase.
u PresstheSTARTbutton.
Thevacuumpumpswitchesonandrunsuntilthesetpressureisreached.There
afterthepumpwillintermittentlystartupautomaticallytomaintainthepressure.
u Ifnecessary,dependingontheventingsetting(seeChapter5.4.1"Controllerset
tings",page20),presstheSTOPbuttontoventthesystem.
Komentarze do niniejszej Instrukcji