6|Operation BÜCHILabortechnikAG
28/42 OperationManualInterfaceI100
6.3.3 Completecalibration(withtemperaturecompensation)
Thiscalibrationmethodisusedtoadjustthelinearityofthepressuresensoratspecific
predefinedpressurelevelsandtemperatures.Duringthecalibrationsequencethe
pressuresensoriscalibratedatroomtemperatureandatatemperatureofapprox.55
°Cinsixstagesatnormalatmosphericpressureandat800,600,400,200and10
mbar.
NOTE
Thiscalibrationprocessmayonlybecarriedoutbyauthorizedservicetechnicians.
Duringthecalibrationsequence,thesystemcarriesoutaplausibilitycheckonthe
measuredreadingsandindicatesafaultiftheconfirmedreadingsareimplausible.
u Connectthereferencepressuregaugetothesamevacuumsystemthatthepump
andinterfaceunitareconnectedto.
u Onthecalibrationmenu,selecttheoptionCompletepressurecalibration.
Fig.19:Menu>Pressuresensorcalibration>Completepressurecalibration.
u Usingthearrowbuttons,adjustthespecifiedreferencepressuretothefigure
shownonthereferencepressuregaugeandpressOKtoconfirm.
u PresstheNextbutton.
u UsingtheP↑andP↓buttons,increase/decreasethevacuuminthesystemuntilthe
pressureshownonthereferencegaugeiswithin±10mbarofthespecifiedsetting.
u Repeatthelastthreestepsforeachofthepressurelevelslistedabove.
u Whenthesixpressurelevelshavebeencalibratedatroomtemperature,placethe
interfaceunitinalaboratorykilnandheatitupto55°C(approx.1.5hours).
u CAUTION!Riskofskinburns.Wearprotectivegloves.Leavetheinterfaceunit
inthekilnandrepeatthestepssetoutabove.
u Oncompletionofthelaststep,pressOKtoconfirmthecalibration.
Thecalibrationsettingsaresavedandthedisplayrevertstothedefaultscreen.